HYDROGEN-ISOTOPE ANALYSIS OF POTENTIALLY TOXIC ORGANIC MATERIALS EMPLOYING MANGANESE REDUCTION AND DISPOSABLE NICKEL-PYROLYSIS TUBES

dc.contributor.authorMotz J.E.
dc.contributor.authorShouakar-Stash O.
dc.contributor.authorDrimmie R.J.
dc.contributor.authorEdwards T.W.D.
dc.contributor.authorFrape S.K.
dc.contributor.authorAnnable W.K.
dc.date.accessioned2021-03-13T07:01:08Z
dc.date.available2021-03-13T07:01:08Z
dc.date.issued2001
dc.description.abstractUse of a simple, disposable apparatus allows the measurement of 2H/1H ratios in small quantities of potentially toxic organic materials without the release of hazardous compounds. The apparatus consists of welded nickel pyrolysis bombs encased in quartz breakseals. Samples are enclosed in the bombs with a manganese-reducing agent. Hydrogen diffuses through the nickel upon heating and is captured in the quartz envelope for analysis. Tests with standard polyethylene foil and cystine, as well as with water and several herbicides, show that this method produces good precision (typically between +/-1.0%% and +/-4.0%%) and is equally applicable to solid and liquid materials.
dc.identifierhttps://www.elibrary.ru/item.asp?id=793157
dc.identifier.citationChemical Geology, 2001, 181, 1-4, 67-71
dc.identifier.issn0009-2541
dc.identifier.urihttps://repository.geologyscience.ru/handle/123456789/26672
dc.subjectHYDROGEN
dc.subjectISOTOPES
dc.subjectORGANIC
dc.subjectTOXIC MATERIALS
dc.subjectWATER
dc.titleHYDROGEN-ISOTOPE ANALYSIS OF POTENTIALLY TOXIC ORGANIC MATERIALS EMPLOYING MANGANESE REDUCTION AND DISPOSABLE NICKEL-PYROLYSIS TUBES
dc.typeСтатья

Файлы